OALogin with OAToolkit does not move the file at the end of an acquisition - WKB50514

Article number: 50514

SYMPTOMS

  • Sample logged into OALogin
  • At the end of the analysis, the data file is not moved to the network location specified in OAToolkit

ENVIRONMENT

  • MassLynx
  • OALogin
  • OAToolkit

CAUSE

The file path to the remote drive destination folder is incorrect.
The file path was determined by browsing through folders in file explorer software, which missed some steps in the folder path.

FIX or WORKAROUND

Manually enter the complete and correct file path to the networked destination folder, not the mapped driver location.