No concentration or percent deviation results in TargetLynx report - WKB31424
Article number: 31424
SYMPTOMS
- No concentration or percent deviation results in TargetLynx dataset
- No calibration information in dataset
ENVIRONMENT
- MassLynx 4.1
- TargetLynx
CAUSE
"Calibrate standards" is not selected in the TargetLynx Create Dataset operations list.
FIX or WORKAROUND
- Under TargetLynx Shortcut, select Process Samples.
- In the Operations list, select Integrate Samples, Calibrate Standards, Quantify Samples.
- Hit OK.
ADDITIONAL INFORMATION
id31424, MLYNX, MLYNXV41, SUPMM, TARLYNX41