Different sample list formats cannot be applied to different MassLynx sample lists in SCNs 1043 or 1045 - WKB275209

Article number: 275209

SYMPTOMS

  • When opening a sample list, the format is not saved with the list as it was previously
  • Cannot link different sample list format files (.fmt) to different sample lists (.spl)
  • A format file can be created and applied to a sample list as normal, but when a different sample list that previously had a different format loaded and saved is then opened, the new sample list has the same format as the previous sample list
  • Different sample list files cannot be saved with different formats
  • This issue affects anyone who uses MassLynx format files
  • SCNs 1043 and 1045 (no problem in SCN 1042 or earlier releases)

ENVIRONMENT

  • MassLynx
  • Any Quad-based MS

CAUSE

Software defect; logged in CRI-6649

FIX or WORKAROUND

This issue is fixed in SCN 1046 and subsequent releases.