Detector does not start acquisition - WKB27258

Article number: 27258

SYMPTOMS

  • Detector does not acquire data
  • Removing the detector from Inlet Configuration and adding it again does not help

ENVIRONMENT

  • MassLynx
  • 2489 UV/Vis Detector
  • 2996 PDA Detector
  • 2998 Photodiode Array Detector
  • ACQUITY UPLC PDA Detector

CAUSE

Events and triggering are not set up correctly. 

FIX or WORKAROUND

  1. Open the inlet method.
  2. Go to Tools > Instrument Configuration.
  3. Click the Events and Triggering button.
  4. Step through the wizard to the Triggering page.
  5. Verify that the correct setting is enabled for triggering the detector: Trigger by Contact Closure or Trigger by Software
  6. Click OK.

ADDITIONAL INFORMATION