Detector does not start acquisition - WKB27258
Article number: 27258
SYMPTOMS
- Detector does not acquire data
- Removing the detector from Inlet Configuration and adding it again does not help
ENVIRONMENT
- MassLynx
- 2489 UV/Vis Detector
- 2996 PDA Detector
- 2998 Photodiode Array Detector
- ACQUITY UPLC PDA Detector
CAUSE
Events and triggering are not set up correctly.
FIX or WORKAROUND
- Open the inlet method.
- Go to Tools > Instrument Configuration.
- Click the Events and Triggering button.
- Step through the wizard to the Triggering page.
- Verify that the correct setting is enabled for triggering the detector: Trigger by Contact Closure or Trigger by Software
- Click OK.
ADDITIONAL INFORMATION