TargetLynx is not integrating one of the compounds - WKB27050

Article number: 27050

SYMPTOMS

  • TargetLynx produces a report that has one compound not integrated
  • The peak of the compound is only partially drawn in the TargetLynx chromatogram pane (for example, the trace stops at 3.2 min, just as the peak starts to form) 
  • Looking at the data in the MassLynx chromatogram, you can clearly see the acquisition range is complete (up to 3.7 min), with the entire peak visible
  • Changing the TargetLynx processing method of the Chromatogram Mass Window from 0.01 to either 0.05 or even 1 Da doesn't make a difference

ENVIRONMENT

  • MassLynx 4.1 SCN 958, SCN 967
  • TargetLynx
  • TOF MRM data

CAUSE

Incorrect processing version

FIX or WORKAROUND

Data must be processed on a MassLynx desktop software version (such as SCN 909), not the acquisition version.

ADDITIONAL INFORMATION

Refer to KCS article WKB554 for the latest version of the MassLynx Desktop.

id27050, MLYNX, MLYNXV41, SCN909, SCN958, SCN967, SUPMM, TARLYNX41