TargetLynx is not integrating one of the compounds - WKB27050
Article number: 27050
SYMPTOMS
- TargetLynx produces a report that has one compound not integrated
- The peak of the compound is only partially drawn in the TargetLynx chromatogram pane (for example, the trace stops at 3.2 min, just as the peak starts to form)
- Looking at the data in the MassLynx chromatogram, you can clearly see the acquisition range is complete (up to 3.7 min), with the entire peak visible
- Changing the TargetLynx processing method of the Chromatogram Mass Window from 0.01 to either 0.05 or even 1 Da doesn't make a difference
ENVIRONMENT
- MassLynx 4.1 SCN 958, SCN 967
- TargetLynx
- TOF MRM data
CAUSE
Incorrect processing version
FIX or WORKAROUND
Data must be processed on a MassLynx desktop software version (such as SCN 909), not the acquisition version.
ADDITIONAL INFORMATION
Refer to KCS article WKB554 for the latest version of the MassLynx Desktop.