MassLynx injects only the first injection and does not move to the second sample - WKB265392
Article number: 265392
SYMPTOMS
- MassLynx starts the first sample and acquires to the end of acquisition time
- Next acquisition does not start
- MassLynx indicates ready but, in the console, LC controls remain in magenta as during acquisition, yet the MS becomes ready for next acquisition
- In the inlet file, Stop Method is not dimmed, indicating that the method is still running
- Almost every analog data channel is selected for acquisition in the inlet file
ENVIRONMENT
- MassLynx 4.2 SCN1028
- Driver Pack 2022 R1 Analytical
CAUSE
The inlet file is corrupted.
FIX or WORKAROUND
Create a new inlet file using the default inlet file from the default project as a template.