MassLynx injects only the first injection and does not move to the second sample - WKB265392

Article number: 265392

SYMPTOMS

  • MassLynx starts the first sample and acquires to the end of acquisition time
  • Next acquisition does not start
  • MassLynx indicates ready but, in the console, LC controls remain in magenta as during acquisition, yet the MS becomes ready for next acquisition
  • In the inlet file, Stop Method is not dimmed, indicating that the method is still running
  • Almost every analog data channel is selected for acquisition in the inlet file 

ENVIRONMENT

  • MassLynx 4.2 SCN1028
  • Driver Pack 2022 R1 Analytical

CAUSE

The inlet file is corrupted.

FIX or WORKAROUND

Create a new inlet file using the default inlet file from the default project as a template.