QuanOptimize does not work; error "Failed to initiate acqusition method creation software" - WKB257874
Article number: 257874
SYMPTOMS
- Start QuanOptimize after configuring the options in "Run QuanOptimize" and pressing Finish
- The system appears to start but then stops with no error messages
- Open Notepad file in View Results and see the following:
Progress: Error
Compound Ion Transition Mode CV CE
<compound name> 1st In Progress: Optimising CV ESP+Notes:
1. Error occurred creating parent stage method for <compound name>
2. Failed to initiate acquisition method creation software
- Seen with SCN 1040, SCN 1042, and SCN 1043
ENVIRONMENT
- MassLynx
- QuanOptimize
- Xevo TQ-S
CAUSE
Bug in QuanOptimize software - reported in CRI-5754
FIX or WORKAROUND
Fixed in SCN 1045 and subsequent releases
ADDITIONAL INFORMATION
id257874, MLYNX, MLYNXV41, SCN1040, SCN1042, SCN1043, SCN1045, SUPMM, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD