LC-MS system loses communication overnight - WKB248286

Article number: 248286

SYMPTOMS

  • Connection to the MS and LC components are lost overnight
  • Errors in ACQUITY Console indicate lost communication for both UPLC and MS
  • Tune page unresponsive, requiring reboot of PC and MS
  • Time stamp for Console errors corresponds to Windows Event System log errors, which include "windows power setting" 

ENVIRONMENT

  • MassLynx
  • ACQUITY UPLC H-Class
  • Xevo TQ-S micro

CAUSE

PC has "sleep after 5 hours" enabled.   

FIX or WORKAROUND

Configure the PC to "never sleep".

ADDITIONAL INFORMATION