"Right valve move h/w fault" error on ACQUITY UPLC Column Manager - WKB245884
Article number: 245884
SYMPTOMS
- "Right valve move h/w fault" error on ACQUITY UPLC Column Manager
ENVIRONMENT
- Empower
- MassLynx
- ACQUITY UPLC Column Manager
CAUSE
CM-A configured incorrectly in Select mode for 2D UPLC system
FIX or WORKAROUND
Change the configuration to Advanced mode.
ADDITIONAL INFORMATION
When returning column manager to a 1D system, reconfigure the CM-A to Column Selection mode.