How to configure MassLynx to allow a third-party inlet system that is not controlled by MassLynx to trigger the mass spectrometer to start acquiring - WKB24586

Article number: 24586

OBJECTIVE or GOAL

Configure MassLynx to allow a third-party inlet system to trigger the mass spectrometer to start acquiring.

ENVIRONMENT

  • All versions of MassLynx 3.x and 4.x
  • All mass spectrometers

PROCEDURE

  1. Open the MassLynx inlet method.
  2. Go to Tools > Instrument Configuration.
  3. Click the Configure button.
  4. Step through the Configuration Wizard, selecting None for pump and autosampler and clearing all detectors, and then click Finish.
  5. Click the Events and Triggering button.
  6. Click Next.
  7. Select Event In 1, click Next, click Next again, and then click Finish.
  8. Connect the event start cable from the third-party inlet to the Event In 1 connection on the back of the mass spectrometer.
  9. The configuration of the contact closures on the mass spectrometer can be found in the MassLynx console in the mass spectrometer section.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.