In MassLynx, the mass spectrometer does not wait for injection before acquiring - WKB235203
Article number: 235203
SYMPTOMS
- Can control the LC via the MS console
- When submitting a sample from the sample list, MassLynx immediately says "Acquiring" and there is no injection or LC status change
- No data is seen acquiring in real-time updates
- Recently switched inlets from GC to LC system
- Event in 1 is configured for trigger by software for pump and detector
- Rebooting system does not change the status
ENVIRONMENT
- MassLynx
- ACQUITY UPLC H-Class
- Agilent 7890 GC
- Xevo TQ-S micro
CAUSE
GC is on and connected.
FIX or WORKAROUND
Turn off the GC.
ADDITIONAL INFORMATION