In MassLynx, the mass spectrometer does not wait for injection before acquiring - WKB235203

Article number: 235203

SYMPTOMS

  • Can control the LC via the MS console
  • When submitting a sample from the sample list, MassLynx immediately says "Acquiring" and there is no injection or LC status change
  • No data is seen acquiring in real-time updates
  • Recently switched inlets from GC to LC system
  • Event in 1 is configured for trigger by software for pump and detector
  • Rebooting system does not change the status

ENVIRONMENT

  • MassLynx 
  • ACQUITY UPLC H-Class
  • Agilent 7890 GC
  • Xevo TQ-S micro

CAUSE

GC is on and connected.

FIX or WORKAROUND

Turn off the GC.

ADDITIONAL INFORMATION

 