Cannot load GC method for APGC - WKB2350
Article number: 2350
SYMPTOMS
- When the GC method loads, the inlet editor shows a red OK light
ENVIRONMENT
- MassLynx 4.1
- Agilent 7890 GC and 7693 Autosampler
- APGC
CAUSE
The method is configured for a back injector, but there is only a front injector present.
FIX or WORKAROUND
- Determine where the injector physically resides on the GC instrument (front or back).
- Modify the method for the correct injector position. Autosampler > Injector > [select front or back], and ensure that the injector type is correct.
- Save the method with a new name.
- Load the method. In the inlet editor, select the GC drop-down and load method.
- Wait for two green lights (for Ready and OK) on the status tab of the inlet editor.
ADDITIONAL INFORMATION
If this does not sort the load method error: Open Default.A7890 and create a new method from scratch using the default method as a template.
