Communication failure error appears in the MassLynx MS console - WKB230228
Article number: 230228
SYMPTOMS
- In the MS console, "Communication failure / Software communication failure" seen when the IntelliStart option is picked for the MS:
- MassLynx's tune page can control the MS
- When exiting out of all programs and running StopAcquityprocess.cmd, followed by software restart, the software does not clear the error
- There is no resolve button in the MS console for the MS to select
- Resetting instruments in the MS console does not fix the issue
- Power cycling the entire system does not fix the issue
ENVIRONMENT
- MassLynx 4.2
- SYNAPT G2-S MS/HDMS
- SYNAPT G2
- ACQUITY UPLC H-Class
CAUSE
Unknown
FIX or WORKAROUND
1. Run the Masslynx Installer and choose the repair function
2. Reboot EPC afterwards
ADDITIONAL INFORMATION
id230228, comms, MLYNX, MLYNXV41, SUPMM, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO