How to convert between M-Class/nano and standard flow ACQUITY - WKB21195

Article number: 21195

OBJECTIVE or GOAL

Configure a MassLynx-based mass spectrometry system to switch between low flow (nanoACQUITY or M-Class) and standard analytical flow (I-Class, H-Class, ACQUITY Classic).

ENVIRONMENT

  • MassLynx
  • ACQUITY H-Class
  • ACQUITY I-Class
  • ACQUITY M-Class
  • ACQUITY Classic
  • nanoACQUITY
  • Waters Pump Control

PROCEDURE

The attached document outlines a detailed procedure for the case of a system with both an I-Class ACQUITY and an M-CLASS set up for dual pump trapping.

ADDITIONAL INFORMATION

The error message below was seen when the user did not have admin right when trying to run the shortcut. Right click and run as administrator solved this.

Application attemted to perform an operation not allowed by the security policy. To grant this application the required permission, contact your system administrator or us ethe microsoft .net framework confiuration tool

id21195, MLYNX, MLYNXV41, SUPMM, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPMASM, UPMBSM, UPMSMFL, UPNAYXPMP, UPNSAMMGR, UPNSOLMGR, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO