Noise-reduced HDMSE data fails to import to Progenesis QI - WKB208082

Article number: 208082

SYMPTOMS

  • Importing HDMSE raw noise-reduced data to Progenesis QI using Waters Compression Tool fails with a generic error code (109, 0x6d).
  • The original raw data (none noise reduced) imports correctly

clipboard_ec977e7f5a23e120038325f472a37e0d5.png

ENVIRONMENT

  • SYNAPT XS
  • MassLynx
  • Progenesis QI
  • Waters Compression Tool (SCN968)

CAUSE

This occurs when flow is diverted to waste at the start of the analysis.

FIX or WORKAROUND

SOLUTION 1

  • Remove the flow diversion to waste.

SOLUTION 2

  • Keep flow to waste for the desired time. Start the MS acquisition after the diverter valve switches to MS.

ADDITIONAL INFORMATION