Sample manager does not make injection - WKB207580
Article number: 207580
SYMPTOMS
- Sample manager does not make injection
- MS shows "Waiting for Inlet"
- MassLynx queue shows "Acquiring"
ENVIRONMENT
- ACQUITY H-Class FTN
- ACQUITY I-Class FL
CAUSE
Error condition on the sample manager, even after reset
FIX or WORKAROUND
- Cycle power to the sample manager.
- Recover any components that lose communication.
- Establish initial conditions/equilibration.
- Start run.
ADDITIONAL INFORMATION
Typically occurs after an SM error such as "Missing Plate or Vial". Reset does not resolve.