MassLynx is sluggish or not responding when creating the sample list - WKB19519
Article number: 19519
SYMPTOMS
- MassLynx is slow or not responding when creating a sample list
- Circle pointer seen on adding samples to the sample list
- Time lags and delays seen in data collection
ENVIRONMENT
- MassLynx 4.1
CAUSE
Too many data files in the data folder for the current product. The operating system must read through the files in order to create a new data file. If there are a lot of files to read through, it bogs down.
FIX or WORKAROUND
Limit the number of data files in the Project.PRO > Data folder to less than 3000 by doing the following:
- Backing up and deleting the data from the data folder
- Creating a new project every month
ADDITIONAL INFORMATION