Desolvation temperature ramps up to 550 even though the tune page is set to 450 - WKB19211

Article number: 19211

SYMPTOMS

  • When running an LC gradient only, the desolvation temperature doesn't change from the set value of 450
  • When running an LC-MS method via the sample list, the tune page changes from 450 to 550 and then, in a few minutes, comes back down to 450
  • The experimental record for the data file shows a desolvation temperature of 550
  • Both the MS method and the tune page match the resolution mode set

ENVIRONMENT

  • MassLynx 4.1 SCN 923
  • Xevo G2-XS

CAUSE

Corrupted tune page file

FIX or WORKAROUND

  1. Create a new MS Tune file from File > New.
  2. Save the Tune page and use in acquisition.

ADDITIONAL INFORMATION

Same problem already seen for the collision energy.