Skip to main content
Waters

TargetLynx method signature time and method file time may be slightly different in MassLynx Security - WKB18061

Article number: 18061

SYMPTOMS

  • There is sometimes a small difference (for example, 1 second) between the signature timestamp that is displayed in the header of the TargetLynx Method report and the timestamp for the method file that is displayed at the top of the TargetLynx Sample Summary Report and other reports

ENVIRONMENT

MassLynx 4.1 Security

CAUSE

This is expected behavior. There is no guarantee that the times are exactly the same, because they reflect different software processes that occur when the method file is created:

  • The signature time is the time that is stored in the MassLynx checksum file.
  • The method file time is the time that is stored in the operating system.

The method file time may sometimes be very slightly later that the signature time, especially if the operating system or hard disk is slow. Both times are correct.

FIX or WORKAROUND

None 

ADDITIONAL INFORMATION

 

id18061, MLYNX, MLYNXV41, SUPMM, TARLYNX41

 

Not able to find a solution? Click here to request help.