MassLynx acquires some runs but not others - WKB13013

Article number: 13013

SYMPTOMS

  • MassLynx stops acquiring at random intervals—sometimes immediately, sometimes in the middle of a sample set
  • The last data file has no data in it

ENVIRONMENT

  • MassLynx 4.1

CAUSE

Illegal characters are used.

FIX or WORKAROUND

  1. Do not use illegal characters: +  -  . ! @ # $ % ^ & * ( ) =
  2. Use file names with underscores instead of spaces.
  3. Keep file names under 22 characters long.

ADDITIONAL INFORMATION