SYMPTOMS
- Project copied from nonregulated to regulated environment in Windows Explorer
- Logged in to MassLynx as regulated user
- Opened project but "Cannot open checksum file" error occurs
ENVIRONMENT
- MassLynx 4.1 with full security
CAUSE
Project must be imported into regulated environment to create checksum files.
FIX or WORKAROUND
- Log in to MassLynx as a regulated user.
Note: Make sure that the project that opens does not display the checksum error. Change to a different regulated project if necessary. - Open Windows Explorer.
- Browse to the nonregulated project, and copy it to the regulated environment.
- Select Security > Import project in MassLynx, and then highlight the project copied in step 3.
- Specify the MassLynx user and the MassLynx administrator in the confirmation window.
Note: You cannot have administrator in the MassLynx user field. - Click OK.
- Open the project.