"Unable to create alias.txt" error seen during MassLynx installation - WKB119803
Article number: 119803
SYMPTOMS
- While the installation wizard is running, the message "unable to create alias.txt" appears
- Clicking OK allows the installation to proceed, but it ultimately fails
ENVIRONMENT
- MassLynx 4.2 SCN 1012
CAUSE
UAC controls are not at the correct level, and/or the user is not installing as administrator.
FIX or WORKAROUND
- Use Windows search to find "change User Account Control settings" and open it.
- Change the slider to "Never Notify" and click OK.
- Restart the workstation.
- Locate the installation setup.exe; right click and select Run as Administrator.
- Confirm that the software installation completes successfully.
ADDITIONAL INFORMATION
id119803, MLYNX, MLYNXV41, SCN1012, SUPMM