MassLynx "server busy" errors - comm timeout issue - WKB119617
Article number: 119617
SYMPTOMS
- Server busy errors randomly appear and may stop a run
ENVIRONMENT
- MassLynx 4.1
- MassLynx 4.2
- Prep System with 2998/QDa
CAUSE
Comm timeout issue
FIX or WORKAROUND
-
Change the default setting at:
HKEY_LOCAL_MACHINE\SOFTWARE\Wow6432Node\Micromass\Inlets\InletKernel\Settings
Increase Comm Timeout default (20000) to 60000.
ADDITIONAL INFORMATION
Reboot the PC for the new setting to take affect.
id119617, 2998, 2998B, 2998SFC, A-10PDA, QDA