Cannot open OALogin User Configuration to edit or add new users, groups, or methods - WKB112473
Article number: 112473
SYMPTOMS
- Unable to edit or add new users, groups, or methods in OALogin User Configuration
- OALogin software freezes with no errors, and the software closes only with the use of Task Manager
- Reloading OALogin does not resolve issue
- Logged in as Administrator with no policies applied
ENVIRONMENT
- MassLynx
- OpenLynx
- OALogin
- FractionLynx
CAUSE
Unsupported antivirus software restricts the permissions of the OALogin software.
Examples:
Webroot SecureAnywhere Antivirus software (even with exceptions per WKB21150)
Sophos Antivirus Software (even with exceptions per WKB21150)
Cortex with inadequate exclusions
Windows Defender
FIX or WORKAROUND
- Remove the unsupported antivirus software.
- It may be possible to configure the AV software to allow the OALogin to operate correctly, but this would take careful configuration by local IT (unsupported by Waters).