ALS error when performing an injection with 7890B GC and 7850 G4567A sampler on Empower with ICF 3.1 - WKB98921

Article number: 98921

SYMPTOMS

  • When performing an injection with a 7890B GC and 7850 G4567A, an ALS error occurs
  • On the diagnostics tab, it states ALS Failure Code 206: Autoinjector front tower syringe error 
  • Creating a new method does not work
  • Reconfiguring instrument using the Pre-Configuration utility does not resolve the issue
  • When injecting using the 7697A Headspace, it works as expected

ENVIRONMENT

  • Empower 3
  • Instrument Control Framework 3.1
  • Agilent 7890B GC
  • G4567A autosampler

CAUSE

The ALS syringe has a travel lock still in place and the syringe cannot move. 

FIX or WORKAROUND

  1. Remove the travel lock.
  2. Perform an injection.

ADDITIONAL INFORMATION