How is an Agilent GC configured in Empower with ICF 3.X? - WKB98861

Article number: 98861

ENVIRONMENT

  • Empower 3
  • Instrument Control Framework 3.X (ICF 3.X)
  • Agilent 6850 GC
  • Agilent 6890N GC
  • Agilent 6890Plus GC
  • Agilent 6890A GC
  • Agilent 7820 GC
  • Agilent 7890A GC
  • Agilent 7890B GC
  • Agilent 7890A+ GC
  • Agilent Intuvo 9000 GC
  • Agilent 8860 GC
  • Agilent 8890 GC
  • Agilent 7697A Headspace
  • Agilent G1888A Headspace

ANSWER

Configure the GC via the Agilent PreConfiguration Utility. See the Release Notes for the ICF you are using.

ADDITIONAL INFORMATION

Ensure that the Agilent GC meets the minimum firmware requirements and has a static IP address configured. 

Agilent 88XX series GCs require at least the Agilent ICF v3.3 software.