Unable to perform injections with Agilent GC, gas sampling valve, and external samplers with ICF 3.0 (or higher) and Empower - WKB98694
Article number: 98694
SYMPTOMS
- Unable to perform injections with Agilent GC, gas sampling valve, and external samplers with ICF 3.0 (or higher)
ENVIRONMENT
- Empower 3 Personal
- Empower 3 Workgroup
- Empower 3 Enterprise
- Agilent Instrument Control Framework 3.0 (or higher)
- Agilent GC
- Agilent Headspace Sampler
- Agilent 6890 GC
- Agilent 7890 GC
- Agilent 8860 GC
- Agilent 8890 GC
CAUSE
If more than one gas sampling valve or external sampler is connected, the first valve is always assigned as the injection source.
FIX or WORKAROUND
- Switch off GC and Headspace sampler.
- Remove tray and tower connection from the back of the GC.
- Switch on GC only.
- From GC front panel in a configuration, select valve 1 as GSV.
- Remove Headspace sampler from Agilent PreConfig Utility.
- Scan GC from Agilent PreConfig Utility.
- Create a new instrument method (injector preference parameter was not available here as configuration contains GC only).
- Start the analysis.