Empower sample set becomes stuck at "Preparing for Injection" - WKB95534
Article number: 95534
SYMPTOMS
- Empower sample set becomes stuck at "Preparing for Injection" and no sample injected
ENVIRONMENT
- Empower 3
- ACQUITY UPLC H-Class
- Alliance HPLC System
- 2695 Separations Module
- 3465 Electrochemical Detector
CAUSE
System is not communicating correctly—most likely a communication issue with the FTN.
FIX or WORKAROUND
- Close all windows in Empower.
- Power-down the modules and the Empower PC.
- Restart the Empower PC and log in to Windows.
- Turn on the modules
- Log in to Empower and launch Run Samples.
- Load a method and run samples.