The Empower report method is not reporting the correct values - WKB94684

Article number: 94684

SYMPTOMS

  • Two analytes are detected on two different channels
  • Report method shows the correct value for one analyte but not the other

ENVIRONMENT

  • Empower 3
  • Empower 3 Personal
  • Empower 3 Workgroup
  • Empower 3 Enterprise

CAUSE

The processing method components tab has channel A for both analytes.

FIX or WORKAROUND

Change one of the analytes to channel B.  

ADDITIONAL INFORMATION

 