Freezing with an MFC application error when opening the editor screen of a report method created with Empower 3 SR3 or earlier - WKB91292
Article number: 91292
SYMPTOMS
- Occasionally appears after upgrading to SR3
- No issue observed with report methods newly created in this version
- If you open the report method created in the previous version by double-clicking from the Method tab, the following error appears:
MFC application has stopped working
The server threw an exception.
This error causes the freezing.
ENVIRONMENT
- Empower 3 SR3
CAUSE
Report method was created in previous version.
FIX or WORKAROUND
Create a new report method in SR3.
ADDITIONAL INFORMATION