In Empower 3 with Agilent ICF support: Monitoring without signal - WKB91006

Article number: 91006

SYMPTOMS

  • When Monitor is selected, the run sample goes into monitoring, but no signal appears on the screen
  • The Agilent LC control panel indicates that all instruments are in Run mode
  • The same issue occurs for an injection
  • This arises after modifying modules (for example, adding a detector).in the Agilent preconfigurator

ENVIRONMENT

  • Empower Enterprise
  • Agilent ICF 3.2
  • Agilent LC

CAUSE

Unknown

FIX or WORKAROUND

  1. From the Agilent Preconfiguration utility (Tools > Agilent PreConfiguration) type the IP address or name of the LAC/E32, and then remove the instrument.
  2. Add the instrument to the Agilent Preconfiguration utility.
  3. Do not modify the system, the node, or the Waters DHCP Server configuration in Empower.