Error: Startup prerequisite test failed on QDa in Empower - WKB90927

Article number: 90927

SYMPTOMS

Error message in console: Startup prerequisites test failed.

ENVIRONMENT

  • ACQUITY QDa Mass Detector
  • Empower 3

CAUSE

Waters File Service was set to Manual, so not starting up automatically.

FIX or WORKAROUND

  1. Change Startup type to Automatic for the WatersFileService.
  2. Start WatersFileService.

ADDITIONAL INFORMATION

 