How to manually push Empower data off a LAC/E32 - WKB88320
Article number: 88320
OBJECTIVE or GOAL
Manually push Empower data off a LAC/E32
ENVIRONMENT
- Empower Enterprise
- LAC/E32
- Empower data sits in a temporary project folder on LAC/E32 due to a buffering event
- Empower data does not naturally move to the Empower Server
PROCEDURE
- Go to Microsoft Services on the LAC/E32 and stop the WatersService.
- Open the CMD window and type in the following commands:
- cd\ <enter>
- cd empower <enter>
- cd bin <enter>
- instrumentserver -recover <enter>. You will see nothing in the CMD window; it will drop back to the prompt.
3. Wait about 10 seconds, and then start the WatersService.
4. You can now check the recover log on the LAC/E to see if the data has recovered.
ADDITIONAL INFORMATION
If other acquisitions on other systems are presently running on the LAC/E32, do not stop the WatersService, just execute the command.