When data buffering occurs in LAC/Es located in Brazil, the acquisition timestamp is off by an hour compared with the US server time - WKB83824
Article number: 83824
SYMPTOMS
- When data buffering occurs in LAC/Es located in Brazil, the acquisition timestamp is off by an hour compared with the US server time
- Issue arose after Brazil stopped participating in daylight savings time
ENVIRONMENT
- Empower 3 Feature Release 4 (FR4)
- Empower 3 Feature Release 5 (FR5)
- Enterprise / Client server
CAUSE
Empower is using an Oracle time zone file that is not at version 34. (Oracle fixed the issue in version 34.)
FIX or WORKAROUND
- If you have your own Oracle license, download and install the correction from article on the Web: Updated DST Transitions and New Time Zones in Oracle RDBMS and OJVM Time Zone File Patches (Doc ID 412160.1, last updated on November 05, 2019).
-
If you do not have your own Oracle license, change the Node properties of the Brazil computer.
-
Waters logged this issue in CRI 1187.
ADDITIONAL INFORMATION
Which Oracle DST Timezone patch versions are used in Empower 3? WKB27844
CRI-1035 was opened to add version 34 to the next release.