"Vial Attribute not valid" error message with Agilent 7890B and 7697A using Empower - WKB8315
Article number: 8315
SYMPTOMS
- When run begins, message center shows vial attribute not valid
- Instrument failure in the Run Samples
ENVIRONMENT
- Agilent GC 7890B
- Agilent Headspace 7697A
- Empower 3 Service Release 2 Hotfix1
CAUSE
The injection volume in the samples table is too high.
FIX or WORKAROUND
- Adjust the injection volume in the sample set method.
- Maximum injection volume should not exceed 50% of the syringe volume.
- Either reduce the injection volume to less than 5 uL or install a larger syringe.
- Define the correct syringe volume by clicking the link in the Run Sample control panel for Fr. Syringe or Bk. Syringe.
ADDITIONAL INFORMATION
See article WKB62833 on Split injection technique
See article WKB62831 on Splitless injection technique
See article WKB62933 on PTV injection technique