"Vial Attribute not valid" error message with Agilent 7890B and 7697A using Empower - WKB8315
Article number: 8315
SYMPTOMS
- When run begins, message center shows vial attribute not valid
- Instrument failure in the Run Samples
ENVIRONMENT
- Agilent GC 7890B
- Agilent Headspace 7697A
- Empower 3 Service Release 2 Hotfix1
CAUSE
The injection volume in the samples table is too high.
FIX or WORKAROUND
- Adjust the injection volume in the sample set method.
- Maximum injection volume should not exceed 50% of the syringe volume.
- Either reduce the injection volume to less than 5 uL or install a larger syringe.
- Define the correct syringe volume by clicking the link in the Run Sample control panel for Fr. Syringe or Bk. Syringe.