Skip to main content
Waters

"Vial Attribute not valid" error message with Agilent 7890B and 7697A using Empower - WKB8315

Article number: 8315

SYMPTOMS

  • When run begins, message center shows vial attribute not valid
  • Instrument failure in the Run Samples

ENVIRONMENT

  • Agilent GC 7890B  
  • Agilent Headspace 7697A
  • Empower 3 Service Release 2 Hotfix1

CAUSE

The injection volume in the samples table is too high.

FIX or WORKAROUND

  1. Adjust the injection volume in the sample set method.
  2. Maximum injection volume should not exceed 50% of the syringe volume.
  3. Either reduce the injection volume to less than 5 uL or install a larger syringe.
  4. Define the correct syringe volume by clicking the link in the Run Sample control panel for Fr. Syringe or Bk. Syringe.

Syringe info _7890.PNG

ADDITIONAL INFORMATION

See article WKB62833 on Split injection technique

See article WKB62831 on Splitless injection technique

See article WKB62933 on PTV injection technique

id8315, 7890 GC, 7890 GCDET, 7890BGC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, HEADSPACE, SUP

Not able to find a solution? Click here to request help.