Error: Manual Resolution is required, as method is inconsistent with the current hardware configurations - WKB83042
Article number: 83042
SYMPTOMS
- Set up GC method for monitoring in Run Sample window
- Error "Manual Resolution is required, as method is inconsistent with the current hardware configurations"
- Method Resolution Report:

ENVIRONMENT
- Empower 3
- Empower 3 Personal
- Empower 3 Workgroup
- Empower 3 Enterprise
- Agilent 7890 GC
- Agilent 7697A Headspace
CAUSE
The Agilent Headspace was not configured in the Preconfiguration. The column follows the inlet and gives the error if there's no inlet configured.
FIX or WORKAROUND
Go to Preconfiguration in Empower Configure the System and configure the Headspace.