Error: Buffering would overwrite - WKB756

Article number: 756

SYMPTOMS

  • Message in Empower Message Center: "Buffering would overwrite".

  • Status in the Run Samples window: "System Idle - Buffering would overwrite".

ENVIRONMENT

  • Empower

CAUSE

  • Error occurs when restarting an interrupted Sample Set.
  • Indicates buffering and that information will be lost if the action requested is completed.
  • "Overwrite" notifies the user that the existing Samsetxxxx.dat file will be replaced (or overwritten) with new information if the action is allowed to proceed.
  • To preserve integrity, Empower is designed to prevent you from performing this kind of action.

FIX or WORKAROUND

Correct the underlying buffering issue. Once the buffered data is uploaded, the Sample Set can be re-injected.