In Empower 3, cannot open or save Tune pages for TQ Detector/SQ Detector/SQD2 - WKB75574
Article number: 75574
SYMPTOMS
- Tune page initially opens to Default, but when selecting File > Open, no dialog box appears
- Tune page is not linked to any user or project
- Unable to save Tune page
ENVIRONMENT
- Empower 3 Personal
- Empower 3 Workgroup
- Empower 3 Enterprise
- SQ Detector (SQD)
- TQ Detector (TQD)
- SQD2
CAUSE
Special characters are present in the user's password.
FIX or WORKAROUND
Change the password to one that uses no special characters.