Misidentified peaks in Empower - WKB74933
Article number: 74933
SYMPTOMS
- Two peaks have very close retention times
- One peak is misidentified as the peak that occurs next to it
- The other peak is identified as an unknown
- Changing retention times and the retention time window do not cause the peaks to be identified properly
ENVIRONMENT
- Empower 3
CAUSE
The Peak Match column on the Components tab of the Processing Method is set to Closest.
FIX or WORKAROUND
Set the Peak Match in the Components tab of the Processing Method to Greatest Height.
ADDITIONAL INFORMATION
id74933, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP