Empower : Missing peak event if RT window is ok and no channel is selected in processing method - WKB72087
Article number: 72087
SYMPTOMS
- Missing peak
- Peak Codes: Q68
- RT window is well-configured
- No channel selected in the processing method
ENVIRONMENT
Empower 3 FR5
CAUSE
Peak selected as RTReference is missing.
FIX or WORKAROUND
Remove RTReference or choose a Found peak for RTReference.
ADDITIONAL INFORMATION
id72087, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP