Empower : Missing peak event if RT window is ok and no channel is selected in processing method - WKB72087
Article number: 72087
SYMPTOMS
- Missing peak
- Peak Codes: Q68
- RT window is well-configured
- No channel selected in the processing method

ENVIRONMENT
Empower 3 FR5
Empower 3
CAUSE
Peak selected as RTReference is missing.
FIX or WORKAROUND
Remove RTReference or choose a Found peak for RTReference.
ADDITIONAL INFORMATION