Empower : Missing peak event if RT window is ok and no channel is selected in processing method - WKB72087

Article number: 72087

SYMPTOMS

  • Missing peak
  • Peak Codes: Q68
  • RT window is well-configured
  • No channel selected in the processing method

    clipboard_e846e242ee2fe2d4b5feef2a3bc4d07d7.png

ENVIRONMENT

Empower 3 FR5

CAUSE

Peak selected as RTReference is missing.

FIX or WORKAROUND

Remove RTReference or choose a Found peak for RTReference.

ADDITIONAL INFORMATION