QDa "preparing for injection" while running a sample set in Empower - WKB69383
Article number: 69383
SYMPTOMS
- QDa is stuck at "preparing for injection" while running a sample set in Empower
- Unable to perform a single injection from the single inject tab
- Creating a new sample set method does not resolve the issue
- Rebooting the system and LAC/E does not resolve the issue
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC PDA Detector
- ACQUITY QDa Mass Detector
- Empower 3 Enterprise
- Citrix
CAUSE
Corrupt instrument method, or faulty LAC/E network card
FIX or WORKAROUND
Create a new instrument method, or replace the LAC/E network card.
Add UV detector to system. Once UV was added injections indexed as normal.