"Data start" function in Empower does not work with PDA and ethernet controlled detectors - WKB60679

Article number: 60679

SYMPTOMS

  • In the sample set method, a "data start" of 2 minutes is specified, but it does not work

ENVIRONMENT

  • 2475 Multi-Wavelength Fluorescence Detector
  • 2489 UV/Vis Detector
  • 2996 PDA Detector
  • 2998 PDA Detector
  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC PDA eλ Detector
  • ACQUITY UPLC Tunable UV Detector
  • Empower 3

CAUSE

Software defect - PCS 47965

FIX or WORKAROUND

Do not use the Data Start function for these detectors.

ADDITIONAL INFORMATION