System name missing from Empower report - WKB60529
Article number: 60529
SYMPTOMS
- The report does not show a system name; header is blank
- The system name field populates for other system names
ENVIRONMENT
- Empower 3
CAUSE
The cell is too small for the system name.
FIX or WORKAROUND
Adjust the size of the cell or shrink the font.
- Bring the Sample Set/Result Set into Preview/Publisher.
- Select the Report Method that has the issue.
- View the header/footer or, if you have the header on the first page, view the first page header.
- Ungroup the header.
- Manually adjust the size of the cell to be larger where the System Name is displayed. Another option is to change the font size.
- Preview the Report to make sure that the size is large enough.
- Regroup the header.
- Save Report Method.