Sample set does not move to second injection when using a 2487 and 2475 in Empower - WKB60050
Article number: 60050
SYMPTOMS
- When running a sample set with an Alliance 2690, 2487 detector, and 2475 detector, the run does not proceed to the second injection.
- When collecting data, they can see trace for the 2487, but not the 2475.
- If running only 2690 with the 2487, the sample set proceeds to the next injection.
- There is an I/O cable attached between the 2690 and 2475.
ENVIRONMENT
- Empower 3
- Alliance HPLC System
- 2695 Separations Module
- 2475 Multi-Wavelength Fluorescence Detector
- 2487 Dual-Wavelength Absorbance Detector
CAUSE
There is a trigger signal issue between the 2690 and 2475.
FIX or WORKAROUND
Confirm that the I/O cable on the back of the 2690 and 2475 is seated properly and is in the correct I/O signal connectors.
ADDITIONAL INFORMATION
With mix-matched connections, such as IEEE and Ethernet, you need an inject start cable between the 2690 and the 2475.