In Empower, cannot copy instrument and method set methods to a new project - WKB58446
Article number: 58446
SYMPTOMS
- When copying instrument or method set methods to a new project using copy to project, the following error message appears: "Error writing a file because The filename, directory name, or volume label syntax is incorrect"
ENVIRONMENT
- Empower 3
- Project window
- Copy to project...
- Copy (to Desktop or other share destination)
CAUSE
The instrument methods and method sets have hidden characters in their names. This can occur when using a non-English keyboard to enter the names.
There is a restricted character in 'Name' field as defined by TN0000412 – Rev. 00 pg 4-7 , see WKB19581
FIX or WORKAROUND
Rewrite the method name in order to copy to project.
ADDITIONAL INFORMATION