Agilent 6890 acquires only one GC channel even when both channels are enabled in Empower - WKB57124
Article number: 57124
SYMPTOMS
- Agilent 6890 with ALS in the front and G1888 in the back
- There are two detectors, both FID, and both inlets are SSL
- The 6890 acquires only one GC channel even when both channels are enabled in Empower
ENVIRONMENT
- Empower 3 Feature Release 1
- Agilent 6890 GC with firmware N.05.05
- Headspace G1888
FIX or WORKAROUND
- Upgrade firmware to meet the recommended minimum (N.05.06) to be used with Empower 3 FR1.
- Upgrade the version of Empower to a minimum of Service Release 2 (SR2), per PCS 49149.
ADDITIONAL INFORMATION
Abstract/Summary from PCS 49149
Starting with Empower 3 SR2, the Agilent 6890 GC will now acquire two GC channels from a single tower injection when both channels are enabled in the instrument method. For single tower sample set methods, there is no longer a requirement to specify the injector as front (F) or back (B).
id57124, 6890A/+GC, 6890N GC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP