After processing a standard, the peak is not showing its amounts in review - WKB5704
Article number: 5704
SYMPTOMS
- After processing, the standard peak does not have any amounts or a calibration curve.
ENVIRONMENT
- Empower
CAUSE
The spelling of the peak name in the amounts table does not match the spelling of the peak name in the components table. The mismatch could be due to spelling or capitalization.
FIX or WORKAROUND
- Correct the spelling in the components table and reprocess the data.
ADDITIONAL INFORMATION
In the components table, "copy component from processing method" or "copy component from method set" can be used to have the spelling of the peak names match the method.