SYMPTOMS

  • Buffering occurs on the LAC/E during daily backup to the Empower file server at multiple sites
  • Brief intervals of accessibility to the Data Drive.
  • High I/O on the Server Data Drive

ENVIRONMENT

  • Empower
  • Empower 2
  • Empower 3

CAUSE

Backups are scheduled to run while data is being acquired.

FIX or WORKAROUND

  1. Schedule backups for off hours when data acquisition is low.
  2. Backup data to a drive other than the raw data drive.

ADDITIONAL INFORMATION

 