Empower data buffering during backup - WKB53961
Article number: 53961
SYMPTOMS
- Buffering occurs on the LAC/E during daily backup to the Empower file server at multiple sites
- Brief intervals of accessibility to the Data Drive.
- High I/O on the Server Data Drive
ENVIRONMENT
- Empower
- Empower 2
- Empower 3
CAUSE
Backups are scheduled to run while data is being acquired.
FIX or WORKAROUND
- Schedule backups for off hours when data acquisition is low.
- Backup data to a drive other than the raw data drive.
ADDITIONAL INFORMATION
id53961, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP