7890A and 7697A instrument method fails to set up in Empower - WKB53399

Article number: 53399

SYMPTOMS

  • "Set up failed" error message when trying to set up an instrument method with 7890A GC and 7697A Headspace
  • IP address configured correctly as static on the front panel of the GC and Headspace
  • IP addresses for GC and Headspace communicate successfully
  • Instrument information configured correctly in DHCP
  • Headspace firmware is A.01.06 and GC firmware is A.01.16

ENVIRONMENT

  • 7890 driver 2.6
  • Headspace control option 3.0
  • Empower 3 SR2 HF2

CAUSE

Unknown.

Rolling firmware of 7890A GC back to A.01.15 resolves the issue.

FIX or WORKAROUND

Move 7890A firmware back to A.01.15.

ADDITIONAL INFORMATION

Firmware A.01.15 was tested with this driver. For additional firmware information, see the 7890 Instrument Control Software Version 2.6 Release Notes.