Skip to main content
Waters

In Empower, cannot view project audit trail entries for previous month - WKB50976

Article number: 50976

SYMPTOMS

  • No audit trail entries for dates prior to current month

ENVIRONMENT

  • Empower
  • Full Audit Trail Project (FAT) 

CAUSE

Max Rows is set to 1000.

FIX or WORKAROUND

Increase the Max Rows setting, or click on the arrow to proceed to the next page of entries.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.